• Automated Pressure Control: 10,000 PSI
• Manual Pressure Control: 22,500 PSI
• Leak Test Mode
• Depressurisation: (venting)
• Cyclic Pressure Loops
• Profile Writing
• Pressure & Temperature Monitoring
• Supervisor Control & Data Acquisition
• Engineering System & Tolerance Set Up
• Alarm Conditions & Diagnostic Review
The parameters below may vary dependent on test component volume.
• Maximum Operating: 1550 bar (22,500 PSI)
• Pressurisation Control Rates: Linear between 1 and 30 Bar/ min
• Pressurisation Ramp Tolerance: To be maintained within +/- 10% (of full scale)
• Depressurisation Control Rates: Linear between 10 and 50 Bar/ min
• Depressurisation Ramp Tolerance: To be maintained within +/- 10% (full scale)
• Dwell Period: 1 minute to 3 months
• Dwell Period Pressure Stability: 0 (zero) bar to +5 bar
Data Logging
• Input of Batch Set-Up Information
• Data Measurements
• Trend Data
• Capture Data
• Store Data
• Print Reports
Pressure Testing Applications
Effectively simulating subsea conditions to ensure the integrity and reliability of valves, seals, seal assembly’s, connectors, hot stabs, HMHB (Hot Make Hot Break) Connectors, pig locks, heat exchangers, subsea technology and components, for FAT or R&D development.